Aim: To evaluate the accuracy and reproducibility of real and virtual occlusal contact points in implant-supported, fixed complete dentures.
Materials and methods: The study included 19 participants using mandibular interim complete-arch fixed prosthesis supported by 3 or 4 implants as opposed to conventional removable complete dentures. At installation, an examiner installed the prostheses and verified the occlusal contact points through 2 methods: recording the real contact points with carbon paper (RC) followed by occlusal photography and intraoral scanning (VC) to record the virtual contact points to obtain a screen print of the software. Then, the two images were randomized to determine the order to be inserted into Microsoft PowerPoint for blind and paired evaluation. The independent variables consisted of the distribution of occlusal contacts points (qualification through pre-defined scores based on the position of the contact points on the surfaces of the teeth) and the reproducibility of the methods by verifying the number of occlusal points. For this, a descriptive analysis was used to evaluate the distribution of occlusal contacts points and the Wilcoxon test for the reproducibility of the occlusal contact points between the methods (p<0.05).
Results: The methods had 100% and 73.6% real and virtual occlusal contact points, respectively, which is considered clinically excellent. There was no significant difference regarding the reproducibility of the methods by the number of occlusal contact points (RC: x̅13.32; VC: x̅13.68; p=0.715).
Conclusion: The use of the tested intraoral scanner can be an easy and fast tool for studying and mapping the occlusion, and storing data for future treatment, with the conventional method being the preferred method for performing the occlusal adjustment.
Schlagwörter: implant-fixed dental prosthesis, dental occlusion, edentulous, jaw, Occlusal adjustment, three-dimensional imaging