DOI: 10.11607/ijp.6166, PubMed-ID: 31283815Seiten: 358-360, Sprache: EnglischVigolo, Paolo / Fonzi, Fulvio / Zaccaria, Massimiliano / Mutinelli, SabrinaPurpose: To assess the quality of precision at the implant-abutment interface of single-tooth restorations with titanium and zirconia abutments with hexagonal external connections produced using a digital protocol.
Materials and Methods: A total of 20 abutments were produced with commercially pure titanium, and 20 abutments with zirconia, both following a necessary digital protocol. Rotational freedom of all the abutments was assessed. All data were analyzed according to procedures established by software package STATA 14.2.
Results: No significant differences relative to rotational freedom emerged between the two groups.
Conclusion: This study showed that both types of abutments constantly demonstrated approximately 2 degrees of rotational freedom between implant and abutment where the hexagonal external connection was concerned.