Seiten: 38-42, Sprache: EnglischAl-Turki, Lulwa Ebrahim E. / Chai, John / Lautenschlager, Eugene P. / Hutten, Mark C.Purpose: The effect of two levels of prosthesis misfit on prosthetic screw stability was evaluated.
Materials and Methods: Two levels of vertical discrepancies-100 and 175 µm-were introduced between an implant-supported complete denture and the terminal abutment. An implant-supported complete denture without vertical discrepancy served as a control. Cyclic load was delivered vertically on the cantilever portion of the prosthesis next to the terminal abutment for 48 hours for each trial. A total of seven sets of new screws were tested for each level of fit.
Results: The results revealed significant prosthetic screw instability at both the 100- and 175-µm levels of discrepancy.
Conclusion: Vertical discrepancies of 100 and 175 µm introduced between an implant-supported fixed complete denture and the terminal abutment resulted in significant prosthetic screw instability.